Laboratoriets måleevne er fastlagt for intervaller af medie- og lufttemperaturer
Best measurement capabilities are determined within intervals of media (tmedia) and ambient (tair) temperatures.
Område A1 (typisk feltkalibrering)
Range A1 (typically used in the field)
tair ≥ -24 °C + tmedia for 0 °C ≤ tmedia ≤ 14 °C
tair ≥ -45 °C + 2,5 * tmedia for 14 °C ≤ tmedia ≤ 25 °C
tair ≥ 10 °C + 1,3 * tmedia for 0 °C ≤ tmedia ≤ 20 °C
tair ≤ 36 °C for 20 °C ≤ tmedia ≤ 25 °C
Område A2 (typisk kalibrering i felt og prøvehal)
Range A2 (typically calibration in the field or in testing facilities)
| tmedia - tair | ≤ 6 °C for 6 °C ≤ tmedia ≤ 18 °C
Område A3 (kalibrering i prøvehal og laboratorium)
Range A3 (calibration in testing facilities or in the laboratory)
15 °C ≤ tair ≤ 28 °C tair for - 5 °C ≤ tmedia ≤ tair+ 2 °C